3 VMA Mode
3.1 Digital Demod Measurement
Remote
Command
:CALCulate:DDEMod:TRACe:SEARch:TIME:OFFSet <time>
:CALCulate:DDEMod:TRACe:SEARch:TIME:OFFSet?
Example
:CALC:DDEM:TRAC:SEAR:TIME:OFFS 10 us
:CALC:DDEM:TRAC:SEAR:TIME:OFFS?
Notes The align Reference, Offset and Interval determine the trace alignment region, and if the region is out of
the acquisition data range, the trace will stop updating, and a trace indicator “DATA?” (which means
“Data is questionable relative to current setup.”) will be displayed at the top right of the graph
Preset 0 s
State Saved Saved in instrument state
Min -1.0e6 s
Max 1.0e6 s
Interval
Specifies the trace alignment interval. It works with Reference and Offset, determine
the trace alignment region.
Vector Modulation Analyzer Mode User's &Programmer's Reference 406