3 VMA Mode
3.5 Channel Power Measurement
Command
[:SENSe]:CHPower:AVERage:COUNt?
Example
:CHP:AVER:COUN 15
:CHP:AVER:COUN?
Preset SA, WLAN: 10
WCDMA, LTEAFDD, LTEATDD, 5G NR, MSR: 200
State Saved Saved in instrument state
Min/Max 1 / 10000
Annotation The average count is displayed in the measurement bar on the front panel display. The annotation
appears in the format n/N where n is the current average and N is the average count
Averaging On/Off
Turns averaging on or off for this measurement.
NOTE
In this measurement, the Average Type is preset to the Log-Pwr Avg (Video)
method. Other averaging methods are not available.
Remote Command
[:SENSe]:CHPower:AVERage[:STATe] ON | OFF | 1 | 0
[:SENSe]:CHPower:AVERage[:STATe]?
Example
:CHP:AVER ON
:CHP:AVER?
Preset
ON
State Saved Yes
Range
ON|OFF
Avg Mode
Allows you to select the type of termination control used for the averaging function.
This determines the averaging action after the specified number of data acquisitions
(average count) is reached. Options are:
–
EXPonential: The measurement averaging continues using the specified
number of averages to compute each exponentially-weighted averaged value.
The average is displayed at the end of each sweep
–
REPeat: The measurement resets the average counter each time the specified
number of averages is reached
Remote
Command
[:SENSe]:CHPower:AVERage:TCONtrol EXPonential | REPeat
[:SENSe]:CHPower:AVERage:TCONtrol?
877 Vector Modulation Analyzer Mode User's &Programmer's Reference