3 VMA Mode
3.1 Digital Demod Measurement
Clock Error
Sets the limit for the symbol clock error measurement pass/fail test.
Remote
Command
:CALCulate:DDEMod:SEGMent1|…|6:LIMit:CERRor[:HERTz] <real>
:CALCulate:DDEMod:SEGMent1|…|6:LIMit:CERRor[:HERTz]?
Example
:CALC:DDEM:SEGM1:LIM:CERR 40.0
Set the clock error limit of segment1 to 40.0 Hz
:CALC:DDEM:SEGM1:LIM:CERR?
Query the clock error limit of segment1
Preset 100.0
State Saved Yes
Min 0 Hz
Max 500 kHz
Backwards
Compatibility
SCPI
:CALCulate:DDEMod:LIMit:CERRor[:HERTz]
RMS Offset EVM
Available only when the modulation format is Offset QPSK or Shaped Offset QPSK.
Sets the limit for the RMS Offset EVM measurement pass/fail test.
Remote
Command
:CALCulate:DDEMod:SEGMent1|…|6:LIMit:OEVM:RMS <real>
:CALCulate:DDEMod:SEGMent1|…|6:LIMit:OEVM:RMS?
Example
:CALC:DDEM:SEGM1:LIM:OEVM:RMS 10.0
Set the RMS Offset EVM limit of segment1 to 10 %
:CALC:DDEM:SEGM1:LIM:OEVM:RMS?
Query the RMS Offset EVM limit of segment1
Preset 35.0
State Saved Yes
Min 0.00
Min/Max 0.00/100.0
Backwards
Compatibility
SCPI
:CALCulate:DDEMod:LIMit:OEVM:RMS
377 Vector Modulation Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals