6 Input/Output
6.5 Corrections
INPut
Correct the port only when the port is used as an Input
OUTPut
Correct the port only when the port is used as an Output
BOTH
Correct the port when the port is used as either an Input or an Output (or both)
A port that is only an Output is always corrected as an output if the Correction is On.
A port that is only an Input is always corrected as an Input if the Correction is On.
For a port that can be either an Input or an Output (or both), the Correction is
determined by the Correction Direction setting. The default is BOTH, which means
that by default a port that can be either an Input or an Output (or both) is corrected
in both directions if the Correction is On.
Remote Command
[:SENSe]:CORRection:CSET[1]|2|…|16:DIRection INPut | OUTPut | BOTH
[:SENSe]:CORRection:CSET[1]|2|…|16:DIRection?
Example
:CORR:CSET2:DIR INP
Dependencies The Correction Direction control only appears when Correction Port selects a port that can either
function as an input or an output (or both simultaneously), such as RFIO HD, RFFD or T/R. If the SCPI
command is sent to any other port, it is accepted but ignored
Preset Not affected by a Preset. Set to BOTH by Restore Input/Output Defaults
State Saved Saved in State
Backwards
Compatibility SCPI
The following SCPI results in the selection of BOTH (included for compatibility with early Multitouch
implementations):
[:SENSe]:CORRection:CSET[1]|2|…|8:DIRection BIDirectiona
included for compatibility with A-models modular products:
[:SENSe]:CORRection:CSET[1]|2|…|8:RF:PORT:RFFD SOURce | ANALyzer | BOTH
[:SENSe]:CORRection:CSET[1]|2|…|8:RF:PORT:RFIO1 SOURce | ANALyzer | BOTH
[:SENSe]:CORRection:CSET[1]|2|…|8:RF:PORT:RFIO2 SOURce | ANALyzer | BOTH
[:SENSe]:CORRection:CSET[1]|2|…|8:RF:PORT:RFIO3 SOURce | ANALyzer | BOTH
[:SENSe]:CORRection:CSET[1]|2|…|8:RF:PORT:RFIO4 SOURce | ANALyzer | BOTH
6.5.5 Edit Correction
Invokes the integrated editing facility for this correction set.When entering the
menu, the editor window turns on, the selected correction is turned On, Apply
Corrections is set to On, the amplitude scale is set to Log, and the Amplitude
Correction (“Ampcor”) trace is displayed. The actual, interpolated correction trace is
shown in green for the selected correction. Note that since the actual interpolated
correction is shown, the correction trace may have some curvature to it. This trace
represents only the correction currently being edited, rather than the total,
accumulated amplitude correction for all amplitude corrections which are currently
on, although the total, accumulated correction for all corrections which are turned
on is still applied to the data traces.
Vector Modulation Analyzer Mode User's &Programmer's Reference 2212