9 Programming the Instrument
9.2 IEEE 488.2 Common Commands
9.2 IEEE 488.2 Common Commands
The instrument supports the following subset of IEEE 488.2 Common Commands, as
defined in Chapter 10 of IEEE Standard 488.2–1992. As indicated below, some of
these commands correspond directly to instrument front-panel functionality, while
others are available only as remote commands.
"*CAL? - Calibration Query" on page 2542 (Align Now All equivalent)
"*CLS - Clear Status" on page 2543
"*ESE - Standard Event Status Enable" on page 2543
"*ESR? - Standard Event Status Register Query" on page 2544
"*IDN? - Identification Query" on page 2544
"*OPC? - Operation Complete" on page 2545
"*OPT? - Query Instrument Options" on page 2546
"*RCL - Recall Instrument State" on page 2546 (Recall State equivalent)
"*RST - Reset" on page 2546 (Mode Preset equivalent)
"*SAV - Save Instrument State" on page 2547 (Save State equivalent)
"*SRE - Service Request Enable" on page 2547
"*STB? - Status Byte Query" on page 2548
"*TRG - Trigger" on page 2548
"*TST? - Self Test Query" on page 2548
"*WAI - Wait-to-Continue" on page 2549
9.2.1 *CAL? - Calibration Query
*CAL? Performs a full alignment and returns a number indicating the success of the
alignment. A zero is returned if the alignment is successful. A one is returned if any
part of the alignment fails. The equivalent SCPI command is :CALibrate[:ALL]?
See "Align Now All" on page 1902
Remote Command
*CAL
Vector Modulation Analyzer Mode User's &Programmer's Reference 2542

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals