9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Note that the Event register is "Clear-on-Read": when any bit is read, it is
automatically cleared.
Questionable Registers
These registers report abnormal operating conditions. The status register hierarchy
is:
The summary outputs from the six QUEStionable:<keyword> detail registers
are inputs to the "Questionable Register" on page 2592
The summary output from the "Questionable Register" on page 2592 is an input
to the Status Byte Register
The summary output from the is an input to the "Operation Register" on page
2585. The inputs to the "Operation Condition Query" on page 2586 Register
indicate the real time state of the instrument. The "Operation Event Query" on
page 2587 Register summary output is an input to the Status Byte Register
Note that, in E4406A only, the "Operation Enable" on page 2587 Register has an
additional function. It is ANDed with the "Operation Condition Query" on page 2586
Register to determine the instrument busy state, which is checked by "*OPC? -
Operation Complete" on page 2545 and "*WAI - Wait-to-Continue" on page 2549 .
If the ANDed result is non-zero, the instrument is considered busy.
9.4.3 Status Register SCPI Commands
Monitoring of instrument conditions is done at the highest level using the following
IEEE 488.2 common commands.
For complete command descriptions, see "IEEE 488.2 Common Commands" on
page 2542. Individual status registers can be set and queried using the commands
described in "Status Subsystem Registers and Commands" on page 2579.
*CLS
Clear Status Clears the status byte by emptying the error queue and clearing all
the event registers
*ESE
*ESE?
Event Status
Enable
Sets and queries the bits in the enable register part of the standard
event status register
*ESR?
Event Status
Register
Queries and clears the event register part of the standard event
status register
*OPC
*OPC?
Operation
Complete
Sets the standard event status register to monitor the completion of
all commands. The query stops any new commands from being
processed until the current processing is complete, then returns a
‘1’
*PSC
*PSC?
Power-on
State Clear
Sets the power-on state so that it clears the service request enable
register and the event status enable register at power on
Vector Modulation Analyzer Mode User's &Programmer's Reference 2574

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals