3 VMA Mode
3.9 SEM Measurement
State Saved Saved in instrument state
Range
TOP|CENTer|BOTTom
Annotation The greater than (>) and less than (<) symbols are displayed on both sides of the graph to indicate the
Reference Position
Backwards
Compatibility
SCPI
:DISPlay:SEMask:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RPOSition
Auto Scaling
Toggles Auto Scaling On or Off.
Remote
Command
:DISPlay:SEMask:WINDow[1]:TRACe:Y[:SCALe]:COUPle 0 | 1 | OFF | ON
:DISPlay:SEMask:WINDow[1]:TRACe:Y[:SCALe]:COUPle?
Example
:DISP:SEM:WIND:TRAC:Y:COUP OFF
:DISP:SEM:WIND:TRAC:Y:COUP?
Couplings When Auto Scaling is ON, and the Restart front-panel key is pressed, this function automatically sets
the scale per division to 10dB and determines the reference values based on the measurement results
When you set the value of either Scale/Div, Ref Value, or Scale Range manually, Auto Scaling automat-
ically changes to OFF
Preset 1
State Saved Saved in instrument state
Range
OFF|ON
Backwards
Compatibility
SCPI
:DISPlay:SEMask:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:COUPle
3.9.13.2 Attenuation
Controls the attenuator functions and interactions between the attenuation system
components.
There are two attenuator configurations in the X-Series. One is a Dual-Attenuator
configuration consisting of a mechanical attenuator and an optional electronic
attenuator. The other configuration uses a single attenuator with combined
mechanical and electronic sections that controls all the attenuation functions.
Different models in the X-Series come with different configurations.
–
See "Dual-Attenuator Configurations" on page 1421
–
See "Single-Attenuator Configuration" on page 1422
Vector Modulation Analyzer Mode User's &Programmer's Reference 1420