3 VMA Mode
3.2 Custom OFDM Measurement
Example
:POW:MW:PATH:AUTO:FULL ON
:POW:MW:PATH:AUTO:FULL?
Dependencies Only appears if Option FBP is installed, and in the following measurements
5GNRMode: Modulation Analysis and IQ Waveform
WLAN Mode: IQ Waveform
Channel Quality Mode: Group Delay and Noise Power Ratio
Preset
OFF
State Saved Saved in instrument state
Software Preselection
Provided in some instruments, either to compensate for issues with provided
hardware preselection or to provide the preselection function when there is no
hardware preselector.
N9041B
Software Preselection compensates for the frequency range limit of the microwave
preselector. Since the microwave preselector only goes up to 50GHz, software
preselection must be used to suppress and separate images above 50GHz. The
specific algorithm used for software preselection is specified by the SW Preselection
Type selection Normal or Advanced.
In N9041B, Software Preselection only applies for frequencies above 50GHz,
therefore it is only used for RF Input 2. Even if turned on, it is not used for other
inputs, and never for frequencies below 50GHz. This is why the label of this control
contains the parenthetical note “(>~50GHz)”.
Note that in N9041B, in Swept SA measurement, Software Preselection works even
if the measurement is using an FFT Sweep Type. In measurements other than Swept
SA, Software Preselection is not used if the measurement is using an FFT sweep
type.
N9042B+V3050A
Software Preselection compensates for the frequency range limit of the microwave
preselector. Since the microwave preselector only goes up to 50GHz, software
preselection must be used to suppress and separate images above 50GHz. The
specific algorithm used for software preselection is specified by the SW Preselection
Type selection Normal or Advanced.
For N9042B+V3050A, Software Preselection only applies for frequencies above
50GHz, therefore it is only used for External RF. Even if it is turned on, it will not be
Vector Modulation Analyzer Mode User's &Programmer's Reference 460

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals