3 VMA Mode
3.3 Monitor Spectrum Measurement
Scale/Div also determines the displayed amplitude range in the log plot graph.
Since there are usually 10 vertical graticule division on the display, the total
amplitude range of the graph is typically 10x this amount. For example, if Scale/Div
is 10 dB, then the total range of the graph is 100 dB.
Remote
Command
:DISPlay:MONitor:WINDow[1]:TRACe:Y[:SCALe]:PDIVision <rel_ampl>
:DISPlay:MONitor:WINDow[1]:TRACe:Y[:SCALe]:PDIVision?
Example
:DISP:MON:WIND:TRAC:Y:PDIV 5.0 dB
:DISP:MON:WIND:TRAC:Y:PDIV?
Couplings Coupled to Scale Range as follows
Scale/Div = Scale Range/10 (number of divisions)
When "Auto Scaling" on page 566 is ON, this value is automatically determined by the measurement
result
When you change a value, Auto Scaling automatically changes to OFF
Preset 10.00 dB / Div
State Saved Saved in instrument state
Min 0.10 dB
Max 20 dB
Annotation <value> dB/ left upper of graph
Backwards
Compatibility
SCPI
:DISPlay:MONitor:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:PDIVision
Scale Range
Sets the Y Axis scale range.
Remote Command
:DISPlay:MONitor:WINDow[1]:TRACe:Y[:SCALe]:RANGe <rel_ampl>
:DISPlay:MONitor:WINDow[1]:TRACe:Y[:SCALe]:RANGe?
Example
:DISP:MON:WIND:TRAC:Y:RANG 100
:DISP:MON:WIND:TRAC:Y:RANG?
Couplings Coupled to Scale/Div as follows
Scale Range = Scale/Div * 10 (number of divisions)
When you change a value, Auto Scaling automatically changes to OFF
Preset 100 dB
State Saved Saved in instrument state
Min 1
Max 200
565 Vector Modulation Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals