EasyManuals Logo

Keysight x-series Programmer's Reference

Keysight x-series
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1018 background imageLoading...
Page #1018 background image
3 VMA Mode
3.6 Occupied BW Measurement
Sweep Time
Video BANDwidth VBW/RBW ratio
Upper and Lower Tone (set to Sense)
Zero span measurement Resolution Bandwidth
Zero span measurement Dwell Time
Harmonics (SA Mode only)
Parameters affected by Auto Couple are:
Resolution Bandwidth
Fundamental Frequency
Dwell Time
Range Table Resolution Bandwidths
Range Table Dwell Times
Meas Preset
Restores all measurement parameters to their default values.
Remote Command
:CONFigure:OBWidth
Example
:CONF:OBW
Max Hold (Remote Command Only)
When ON, Max Hold displays and holds the maximum responses of the current
measurement. Turn Max HoldOFF to disable the maximum hold feature.
Remote
Command
[:SENSe]:OBWidth:MAXHold ON | OFF | 1 | 0
[:SENSe]:OBWidth:MAXHold?
Example
:OBW:MAXH ON
:OBW:MAXH?
Couplings Max Hold is coupled to "Averaging On/Off" on page 1012. Max Hold is activated only if Average state is
ON. If Max Hold is changed to ON when Average state is OFF, Average state is automatically set to ON
Preset
OFF
Vector Modulation Analyzer Mode User's &Programmer's Reference 1018

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals