EasyManua.ls Logo

R&S ZNB - Page 1010

R&S ZNB
1496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
1010User Manual 1173.9163.02 ─ 55
Parameters:
<Scope> GLOBal | CHANnel
CHANnel when all the sweeps for each channel are complete
GLOBal when all sweeps in all channels are complete
Note:*RST or "Preset" do not change the configuration of the
Universal Interface. Use CONTrol:HANDler:RESet to restore
default values.
*RST: n/a (default: GLOBal)
Example: See CONTrol:HANDler:PASSfail:MODE
CONTrol:HANDler:PASSfail:STATus?
Returns the global pass/fail status of the last measurement.
Return values:
<Status> PASS | FAIL | NONE
PASS all measurements that are not in single sweep mode (on
hold) have been swept, and all limit checks have been passed.
FAIL all measurements that are not in single sweep mode (on
hold) have been swept, at least one limit check failed according
to the specified pass/fail policy (CONTrol:HANDler:
PASSfail:POLicy).
NONE no pass/fail status available, e.g. because the measure-
ment is in progress or because no limit check has been defined.
*RST: n/a
Example: Preparations: Configure and enable a limit check. Start a mea-
surement and wait until the sweep is complete.
CONTrol:HANDler:PASSfail:STATus?
Query the result of the global limit check.
Usage: Query only
CONTrol:HANDler:RESet
Restores the default states of the CONTrol:HANDler... commands including the
data port values.
Example:
See CONTrol:HANDler[:EXTension]:INDex:STATe
Usage: Event
CONTrol:HANDler:SWEepend <SweepEnd>
Specifies the event that will cause the /SWEEP END line (pin 34) to go low; see Chap-
ter 11.3.4.4, "Timing of Control Signals", on page 1438.