Command Reference
R&S
®
ZNB/ZNBT
1327User Manual 1173.9163.02 ─ 55
SYSTem:USER:DISPlay:TITLe......................................................................................1341
SYSTem:USER:KEY.................................................................................................... 1342
SYSTem:VERSion?......................................................................................................1342
SYSTem:AGC:GLOBal:METHod <Global AGC Method>
Allows you to select between two AGC algorithms.
Setting parameters:
<Global AGC
Method>
POINt | MEMory
POINt
Point AGC
MEMory
Memory AGC
Manual operation: See "Global AGC (R&S ZNB only!)" on page 709
SYSTem:CORRection:WIZard[:IMMediate]
<Dialogs>
Keysight-compatible command to open the Start Cal Tab or the Calibration Kits Dialog.
Tip: When working with the Chapter 7.1.3, "GPIB Explorer", on page 743, switch to
raw mode ("Options">"Raw mode" in the IECWIN32 GUI) before executing this com-
mand.
Setting parameters:
<Dialogs> MAIN | CKIT
MAIN: open the "Calibration > Start Cal" softtool tab
CKIT: open the "Calibration Kits" dialog
Usage: Setting only
SYSTem:DATE <Year>, <Month>, <Day>
The command queries or defines the instrument's current date setting.
The setting command requires administrator rights; refer to Chapter 11.1.1, "Windows
Operating System", on page 1423.
Parameters:
<Year> Year, four-digit number
<Month> Month, two-digit number, 01 (for January) to 12 (for December)
<Day> Day, two-digit number, 01 to the number of days in the month
Example:
SYST:DATE?
Response: 2012,05,01 - it is the 1st of May, 2012.
SYSTem:DFPRint?
Queries the device footprint. The device footprint contains detailed information about
the instrument and is mostly used for service purposes.
SCPI Command Reference