EasyManua.ls Logo

Trimble S - Page 409

Default Icon
596 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 6
Adjustment and verification
P/N 57150002, Revision 5.0 6 - 73 Trimble S, VX, SPS & RTS Service Manual
F
4 Adjust reference fibre. A. Adjust the reference
greywedge within
tolerance, see Fig. 6-
77 and Fig. 6-79.
B. Click
Adjust OK button
when finished, see Fig.
6-79.
156000 -
164000, see Fig.
6-79.
There is a five min.
time to adjust the
reference signal after
the warm-up. If more
than five min. has
passed click
Start
Warm-up
button and
another five min.
adjustment window
will start.
5 Remarks The reference greywedge has a “0” position where the area of highest
density jumps to the area of lowest density Fig. 6-78.
The reference greywedge must not be positioned in this high-low
transition area. The adjustment will be very unstable, the wrong
position is characterised by the running bar is reacting much more
sensitive.
The proper adjustment position will allow a larger turn.
Fig. 6-77 Reference greywedge adjustment
Fig. 6-78 Reference greywedge disc
Item Performance Description Result Notes
greywedge
Reference
A
Forbidden range
high
low
density
density
“0” position
Path of the reference light

Table of Contents

Related product manuals