EasyManuals Logo

Keysight x-series Programmer's Reference

Keysight x-series
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #2418 background imageLoading...
Page #2418 background image
7 Save/Recall/Print
7.3 Save
of the captured image is displayed, with the note “This is the image that will be
saved” below it.
After you have completed the save, a message “File image.png saved” (assuming
image.png was the filename you used).
NOTE
As of firmware release A.17.50, sending *CLS (Clear Status) removes any
message displayed on the screen. If you do not want to see the “File saved”
message after sending :MMEM:STOR:SCR (described below), send the following
sequence (substituting your file name for filename.png): :MMEM:STOR:SCR
“filename.png”;*CLS
NOTE
As of firmware release A.19.50, saving a screen image removes any
informational message displayed on the screen before it captures the screen.
This is useful if you are sending “save image” commands in rapid sequence, as it
keeps the “File saved” message from one screen capture from appearing in the
next screen capture. Error messages are still captured.
If you send a succession of screen image commands
too
rapidly, the system may
not have time to remove the previous message before the next screen capture.
Sending screen image commands more rapidly than twice per second is not
advised.
The default path for State Files is:
My Documents\<mode name>\screen
where <mode name> is the parameter used to select the mode with :INST:SEL, for
example, SA for Spectrum Analyzer Mode.
Screen Image files have extension .png. The default filename is Screen_0000.png,
where the 4-digit number is the lowest number that does not conflict with any
filename in the current directory.
After you have completed the save, the Quick Save front-panel key lets you quickly
repeat the last save performed, using an auto-named file, with new current screen
data.
Remote
Command
:MMEMory:STORe:SCReen <filename>
Example Store the current screen image in the file MyScreenFile.png in the default directory:
:MMEM:STOR:SCR "myScreen.png"
Backwards
Compatibility
SCPI
:HCOPy:SDUMp:DATA?
returns the screen image in a <DEFINITE LENGTH ARBITRARY RESPONSE DATA> element.
The response data is IEEE Block format; the controlling computer can strip the header and store the
result as a .png file
Blocking Screen Capture (Remote Command Only)
2418 Vector Modulation Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals