3 VMA Mode
3.5 Channel Power Measurement
[:SENSe]:CHPower:FREQuency:SYNThesis[:STATe]?
Example
:CHP:FREQ:SYNT 1
:CHP:FREQ:SYNT?
Dependencies Does not appear in all models. For models in which the control is not displayed, the SCPI command is
accepted for compatibility, although no action is taken
Preset 3
State Saved Saved in instrument state
Range See "Ranges" on page 893 below
Auto Function
Remote
Command
[:SENSe]:CHPower:FREQuency:SYNThesis:AUTO[:STATe] OFF | ON | 0 | 1
[:SENSe]:CHPower:FREQuency:SYNThesis:AUTO[:STATe]?
Example
:CHP:FREQ:SYNT:AUTO 1
:CHP:FREQ:SYNT:AUTO?
Preset
OFF
Parameter Options, Installed Options & Ranges
The Phase Noise Optimization control lets you optimize the setup and behavior of
the Local Oscillator (LO) depending on your specific measurement conditions. You
may wish to trade off noise and speed, for example, to make a measurement faster
without regard to noise or with optimum noise characteristics without regard to
speed.
Parameter Values Summary
Option # Description
"Balanced" on
page 889
1
–
In instruments with EP0, balances close-in phase noise with spur
avoidance
–
In instruments without EP0 optimizes phase noise for small
frequency offsets from the carrier
"Best Wide-
offset" on page
890
2 Optimizes phase noise for wide frequency offsets from the carrier
"Fast Tuning"
on page 890
3 Optimizes LO for tuning speed
"Best Close-in"
on page 889
4 or 1*
–
In instruments with EP0, emphasizes close-in phase noise
performance without regard to spur avoidance
–
In instruments without EP0, this setting is accepted but no action is
taken
"Best Spurs" 5
–
In instruments with EP0, emphasizes spur avoidance over close-in
Vector Modulation Analyzer Mode User's &Programmer's Reference 888