Chapter 6
Adjustment and verification
P/N 57150002, Revision 5.0 6 - 185 Trimble S, VX, SPS & RTS Service Manual
F
7 Align Tracker Long
Range.
A. Cover then the
collimator complete
with filter, see Fig. 6-
189 on page 6-185 .
B. Select
TLR Adjust
tracker transmitter
check box. and click
Transmitter ON, see
Fig. 6-190 on page 6-
186 .
C. Align the TLR
transmitter to the
collimators cross-hairs.
Turn the two position
screws 1&2 clockwise
or counterclockwise.
D. Lock the position
screws with Glue MS
1855.
< 1 interval.
Fig. 6-189 Adjust TLR module
Item Performance Description Result Notes
Screw2
Screw1
A
C
C