E5071C
1360
Hides the UserForm (object name: frmSingMeas) from the screen.
Line 130
Displays 9 channel windows.
Line 140
Sets the trigger source to "bus".
Lines 160 to 240
These lines turn on or off Continuous Activation mode for each channel
depending on whether the corresponding option buttons are on or off. By
default, the mode is turned on for channel 1 only.
Line 260
Triggers the instrument to start a measurement cycle.
Line 270
Executes the SCPI.IEEE4882.OPC object to suspend the program until the
value of 1 is returned indicating the end of measurement.
Line 290
Displays a measurement completion message.
Line 310
Displays the UserForm (object name: frmSingMeas) on the screen.
Using the SCPI.TRIGger.SEQuence.SINGle object to suspend the
program until the end of measurement (object name:frmSingMeas)
10| Private Sub cmdExit_Click()
20|
30| Unload Me
40|
50| End Sub
60|
70| Private Sub cmdMeas_Click()
80|
90| Dim Dmy As Long
100|

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals