Measurement
309
2. Click Limit Test.
3. Click Fail Sign. Each press toggles between on/off.
Defining Limit Line
To use the limit test, you must first define the limit line. You can define a
limit table for each trace, and you can define up to 100 limit lines
(segments) in a limit table.
Defining a segment
The following steps describe how to define a segment.
1. Press Channel Next/Channel Prev keys and Trace Next/Trace prev keys
to select the trace on which the limit test function will be used.
2. Press Analysis key to display the Analysis menu.
3. Click Limit Test to display the softkeys associated with the limit test.
4. Click Edit Limit Line to display the limit table.
5. Using the limit table, create/edit a segment. Initially, no segments
are entered in the limit table. At the same time, the Edit Limit Line
menu used to create/edit the limit table is displayed.
6. Click Add to add a segment to the limit table and then specify the
segment parameter values shown below.
Segment
Parameter
Description
Type
Select the type of segment from the
following:
OFF Segment not used for the limit test
MIN
Segment at which the minimum is
specified

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals