Product Information
853
measurements are less important for verifying system
performance. Therefore, the limit lines will not appear on the
printouts.
Equipment Used in the System Verification
Supported Verification Kit:
• 85053B, 3.5 mm, 300 kHz - 26.5 GHz
• 85055A, Type-N, 300 kHz - 18.0 GHz
• 85051B, 7 mm, 300 kHz - 18 GHz
The combination of Calibration Kit and Verification Kit
The combination of Calibration Kit and Verification Kit that can be used is
as follows:
Cal kit
Connec
tor
Kit
Type
Verificat
ion Kit
Minimu
m
Freque
ncy
Maximu
m
Freque
ncy
85033
D
3.5 mm
Mechani
cal
85053B - 6 GHz
85033
E
3.5 mm
Mechani
cal
85053B - 9 GHz
85052
D
3.5 mm
Mechani
cal
85053B -
26.5
GHz
85052
C
3.5 mm
Mechani
cal
85053B -
26.5
GHz
85093
B
3.5 mm ECal** 85053B 300 kHz 9 GHz
85093
C
3.5 mm ECal** 85053B 300 kHz 9 GHz
N4433
A
3.5 mm ECal** 85053B 300 kHz 20 GHz
N4691
A
3.5 mm ECal** 85053B 10 MHz
26.5
GHz
N4691
3.5 mm ECal** 85053B 300 kHz
26.6

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals