E5071C
356
setting (ON, OFF, or --)
ON = Error correction is set to ON
OFF =Error correction is set to OFF
-- = Calibration is not performed
PORT-w to
PORT-z:
Single-end port impedance setting when saving the
data. This is not valid when reading the file. When
reading, the reference impedance value Z0 at option
line is always used.
Option Line
The option line is formatted as follows.
# Hz S FMT R Z0
Parameter Description
FMT Data format
RI = Real number- imaginary number
MA = Linear magnitude - angle
DB = Log magnitude - angle
Z0 Reference impedance value
The structure of the data part depends on the combination of the selected
file type and specified port.
When the S parameter of a port on which calibration
is not performed is specified, if data measured with the S
parameter exists, that data are output to a Touchstone file. If
no measurement data exists, 0 (for log magnitude - angle, log
magnitude = -200 dB) is output in the corresponding field.
The following figures show the data structures of files saved in Touchstone
format.
1-port Touchstone file