Measurement
215
6. Enter a file name in the File Name box and press the Save button to
save the loss compensation table.
Recalling loss compensation table
By recalling a loss compensation table saved in the CSV format according
to the Saving loss compensation table, you can set this loss compensation
table.
1. It is possible to recall a file from a different channel/port from
where it was saved.
1. Press Cal key.
2. Click Power Calibration.
3. Select a port (refer to Selecting target port of error correction).
4. Click Loss Compen.
5. Click Import from CSV File to open the Open dialog box. At this time,
CSV File (extension*.csv) is selected as the file type.
6. Select the CSV format file you want to import and press the Open
button to recall the loss compensation table.
7. This operation is not guaranteed under the following two
conditions: 1) you imported a CSV format file created/edited on
a spreadsheet program, or 2) you imported a CSV format file
that had been exported according to Saving loss compensation
table but then modified.
Setting a tolerance for power calibration
The E5071C provides an option to set the tolerance value for power
calibration. This function is to save time for the power calibration by
aborting the calibration process once the measurement value goes outside
the specified tolerance and is NOT to settle the power level within the
specified tolerance. If the power calibration is aborted, the power-level
error-correction function is not turned on.
Setting the tolerance
Follow the steps below to set the tolerance for power calibration:
1. Press Cal key.
2. Click Power Calibration > Tolerance.
3. Enter the tolerance value for power calibration.
Measuring Calibration Data
• Before measuring calibration data, you need to execute the zero
adjustment and calibration of the power sensor. For information