21. For example, standard 1 for the 85033E 3.5-mm calibration kit is a
SHORT standard.
Standard type
The type of standard used to classify a standard model based on its form
and construction. Six standard types are available: SHORT, OPEN, LOAD,
delay/THRU, Unknown Through, and arbitrary impedance.
Standard coefficient
The numeric characteristics of the standard used in the selected model. For
example, the offset delay (32 ps) of the SHORT standard in the 3.5-mm
calibration kit is a standard coefficient.
Standard class
A group of standards used in a calibration process. For each class, the user
must select the standards to use from the 21 available standards.
Subclass
You can register up to 8 standard types. This capability lets you specify a
different standard for each frequency range. You must assign standards to
subclasses you use.
Defining Parameters for Standards
The following figures show the parameters used in defining standards.
Reflection Standard Model (SHORT, OPEN, or LOAD)

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals