EasyManuals Logo

Agilent Technologies E5071C User Manual

Agilent Technologies E5071C
2685 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #132 background imageLoading...
Page #132 background image
Measurement
131
THRU Response Calibration (transmission test)
• Overview
• Procedure
Other topics about Basic Calibration
Overview
In THRU response calibration, calibration data are measured by connecting
a THRU standard to the desired test port. This calibration effectively
eliminates the frequency response transmission tracking error from the
test setup in a transmission test using that port.
2-Port error model (THRU response)
It is also possible to carry out an isolation calibration using a LOAD
standard in the process of THRU response calibration. An isolation
calibration will eliminate isolation error (crosstalk error) from the test
setup in a transmission test using that port.
2-Port Error model (THRU response + isolation)
1. System Z0 should be changed to 1 ohm before calibration when
using waveguide calibration kit and measuring waveguide
devices.
Procedure
1. Press Channel Next/Channel Prev keys to select the channel for which
you want to perform the calibration.
2. Press Cal key.
3. Click Calibrate > Response (Thru) > Select Ports.

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals