118
Calibration Types and Characteristics
The table shows the different types of calibrations and the features of each
method.
Calibra
tion
Metho
d
Standard (s)
Used
Corrected Error Factor
Measurement
Parameters
Characteristic
s
No
calibra
tion
None None All parameters
Low accuracy
Calibration
not required
Respo
nse
Calibra
tion
OPEN or
SHORT
LOAD
Following 2 error terms:
Reflection Tracking (Er)
Directivity (Ed)
S11(Reflection
characteristics at 1
port)
Medium-level
accuracy
Quick
calibration
Isolation
calibration
improves the
accuracy in a
reflection
measurement
of a DUT with
high return
loss
THRU
LOAD
Following 2 error terms:
Transmission Tracking
(Et)
Isolation (Ex)
S21
(1 direction
transmission
characteristics at 2
ports)
Medium-level
accuracy
Quick
calibration
Isolation
calibration
improves the
accuracy in a
transmission
measurement
of a device
with high
insertion loss
1-Port
Calibra
tion
ECal module
(2-port/4-
port)
Following 3 error terms:
Directivity (Ed)
Source Match (Es)
Reflection Tracking (Er)
S11
(Reflection
characteristics at 1
port)
1-port
measurement
with the
highest
degree of
accuracy
Quick
calibration
with low

Table of Contents