Clearing Calibration Data and Calibration Coefficients
Clearing Calibration Data
When the frequency offset function has been disabled
You can use the following command to clear the measurement values of
calibration data executed with :SENS{1-160}:CORR:COLL:OPEN
command, etc.
:SENS{1-160}:CORR:COLL:CLE
When the frequency offset function has been enabled
You can use the following command to clear the measurement values of
calibration data executed with :SENS{1-160}:CORR:OFFS:COLL:OPEN
command, etc.
:SENS{1-160}:CORR:OFFS:COLL:CLE
These clear functions make the temporarily settings during the calibration,
such as trace number and measurement parameters, recover to the
original state.
Clearing Calibration Coefficients
When the frequency offset function has been disabled
You can use the following command to clear the calibration coefficients
used.
:SENS{1-160}:CORR:CLE
When the frequency offset function has been enabled
You can use the following command to clear the calibration coefficients
used.
:SENS{1-160}:CORR:OFFS:CLE
This command does not clear calibration coefficients related to normal
calibration.

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals