EasyManua.ls Logo

Agilent Technologies E5071C

Agilent Technologies E5071C
2685 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E5071C
1020
ES ER ED EL ET EX
Response calibration (OPEN) * *
Response calibration (SHORT) * *
Response calibration (THRU) * *
Enhanced response calibration * * * * *
1-port calibration * * *
Full 2-port calibration * * * * * *
Full 3-port calibration * * * * * *
Full 4-port calibration * * * * * *
2-Port TRL calibration * * * * *
3-Port TRL calibration * * * * *
4-Port TRL calibration * * * * *
If either an invalid calibration coefficient is specified for the
writing command or a nonexistent calibration coefficient is
specified for its reading command, the following error will occur:
23, Specified error term does not exist
Procedures for Writing Calibration Coefficient
You must follow the steps below to write the calibration coefficient.
1. Declare the calibration type to write.
Execute :SENS{1-160}:CORR:COEF:METH:xxxx command
2. Write any calibration coefficient.
Execute :SENS{1-160}:CORR:COEF command as needed for the written
calibration coefficients
3. Validate the calibration coefficients.
Execute :SENS{1-160}:CORR:COEF:SAVE command
Do not execute any other command while writing the calibration
coefficients.This may cause the system to function incorrectly.
If the fixture simulator is turned on, the calibration coefficient
writing will not function correctly. Make sure to turn off the
fixture simulator before execution.

Table of Contents

Other manuals for Agilent Technologies E5071C

Related product manuals