384
Frequency ranges (segments) from the figure above and their
measurement conditions
Start
frequen
cy
Stop
frequen
cy
Numb
er of
points
IF
Bandwidt
h
Swee
p
mode
Stat
e
A 440 MHz 915 MHz 50 50 kHz
Steppe
d
ON
B 915 MHz 980 MHz 130 70 kHz
Steppe
d
ON
C 980 MHz
1.035
GHz
60 50 kHz
Steppe
d
ON
E 1.07 GHz 2 GHz 100 70 kHz Swept ON
G 2.6 GHz 3 GHz 40 70 kHz Swept ON
Conditions for Setting Segment Sweep
The following conditions apply when setting up a segment sweep.
• The frequency range of a segment must not overlap with that of
another segment. (The start frequency of a segment must be higher
than the stop frequency of the immediately preceding segment).