Setting Averaging Trigger Function
• Overview
• Averaging Trigger Function
Other topics about Making Measurement
Overview
The averaging trigger function is used to execute the sweep the number of
times specified by the averaging factor with a single trigger when the
sweep averaging function is ON.
Averaging
Trigger
Function
ON
Performs the sweep the number of times specified by
the averaging factor with a single trigger.
OFF
Performs the sweep once with a single trigger.
The averaging factor is cleared before the start of measurement.
• When the point trigger function is on, its setting has priority,
and you need to generate triggers based on "(number of
measurement points)× (averaging factor)".
• When the sweep averaging function is off, sweep is performed
only once even if the averaging trigger function is set to on.
• The averaging trigger function is valid for all channels. Note that
you can set the sweep averaging function for each channel.
Averaging Trigger Function
1. Setting Averaging Trigger Function
When the sweep averaging function is ON, follow these steps to set the
averaging trigger function.
1. Press Avg key.
2. Click Avg Trigger.
3. Click ON to activate the averaging trigger.
2. Executing Averaging Measurement
1. Press Trigger key.
2. Click Single. The averaging factor is cleared before the start of
measurement, the sweep is executed the number of times specified
by the averaging factor, and then the instrument waits for the next
trigger.