E5071C
434
Measuring Parameters with Cable
• Overview
• Procedure
Other Measurement Examples
Overview
This section introduces an example of how to detect the location of a
mismatch that occurs in a cable by using the time domain function.
Procedure
In this example, a DUT is evaluated according to the steps.
Step Description
1. Setting the Measurement
Conditions
Set the measurement
conditions.
2. Executing Calibration Execute calibration.
3. Connecting the DUT Connect the DUT.
4. Auto Scale Execute auto scale.
5. Setting the Time Domain Function Set the time domain function.
1. Setting the Measurement Conditions
Follow these steps to set the measurement conditions:
Setting Description Key Operation
Presetting Preset > OK
Stop frequency: 3 GHz Stop > 3 > G/n
Number of points: 201 Sweep Setup > Points > 2 > 0 > 1
> x1
Specifying the low-pass mode
sweep condition
Analysis > Transform > Set Freq
Low Pass
Measurement parameter: S11 Meas > S11
2. Executing Calibration