E5071C
856
Flex Factor determines how much of the cable phase uncertainty to include
in determining the limit lines.
Set to 0% (zero) if the cables are held down in a fixture and are not
allowed to move during the calibration and verification.
Set to 100% if the cables are allowed to move a lot.
Calibration Kit Substitution
Non-Agilent calibration kits are not recommended nor supported.
Precautions for Handling Airlines
When you are using the airlines in the Verification Kit, observe the
following practices to ensure good measurement techniques.
Be very careful not to drop the airline's center or outer conductor. Damage
will result if these devices are dropped.
Use proper Electro-Static Discharge (ESD) procedures.
Clean your hands or wear gloves as skin oils will cause a change in
electrical performance.