Setting Control Functions
733
Calibration of the Touch Screen
When you have executed system recovery on the E5071C, you have to
calibrate the touch screen. Follow the procedure described below to
calibrate the touch screen.
1. Press System Key.
2. Click Service Menu > Test Menu > Adjust Touch Screen.
3. The touch screen calibration screen appears.
4. Touch the x mark on the upper left with your finger. The mark x
appears also on the lower left, upper right, and lower right. Touch
the x marks in that order with your finger.
5. Touching the four locations described above with your finger
automatically concludes the touch screen calibration.
6. With no operation on the touch screen calibration screen for a
preset time, it automatically closes and the previous
measurement screen reappears.