EasyManua.ls Logo

Agilent Technologies E5071C

Agilent Technologies E5071C
2685 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Setting Control Functions
739
2. Mode 2
Setting procedure
Follow these steps to set the external test set mode.
1. Press System > Service Menu > External Test Set.
2. Select your desired mode.
3. When the setting of the external test set mode is changed, the
calibration coefficients acquired using the calibration kit or ECal
are cleared, and the error correction function is turned off.
Calibration and Measurement Parameters
When the external test set mode (Mode 1 or Mode 2) is used with the
E5071C, only limited types of calibrations and measurement parameters
are available.
Calibration when using the external test set
When the external test set mode (Mode 1 or Mode 2) is used with the
E5071C, the following calibrations are available. When other calibration
methods are used, correct calibration is not performed.
Open response calibration
Short response calibration
Thru response calibration
Enhanced response calibration
1-port calibration
When ECal is used with Mode 1 or Mode 2 selected, the
connection of test port 4 cannot be detected automatically.
When you perform calibration using the 4-port ECal, set the
connection of port 4 as follows depending on the connection of
test port 1/2/3.

Table of Contents

Other manuals for Agilent Technologies E5071C

Related product manuals