Setting Control Functions
739
2. Mode 2
Setting procedure
Follow these steps to set the external test set mode.
1. Press System > Service Menu > External Test Set.
2. Select your desired mode.
3. When the setting of the external test set mode is changed, the
calibration coefficients acquired using the calibration kit or ECal
are cleared, and the error correction function is turned off.
Calibration and Measurement Parameters
When the external test set mode (Mode 1 or Mode 2) is used with the
E5071C, only limited types of calibrations and measurement parameters
are available.
Calibration when using the external test set
When the external test set mode (Mode 1 or Mode 2) is used with the
E5071C, the following calibrations are available. When other calibration
methods are used, correct calibration is not performed.
• Open response calibration
• Short response calibration
• Thru response calibration
• Enhanced response calibration
• 1-port calibration
• When ECal is used with Mode 1 or Mode 2 selected, the
connection of test port 4 cannot be detected automatically.
When you perform calibration using the 4-port ECal, set the
connection of port 4 as follows depending on the connection of
test port 1/2/3.

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals