E5071C
324
3. Specify any folder in which you want to save the file, and enter the
file name. Click Save to save the ripple limit table displayed on the
screen to a file.
4. To recall the saved ripple limit table, click Import from CSV File from
the Edit RippleLimit menu to display the Open dialog box. At this
time, CSV Files (with the extension *.csv) is selected as the file type.
5. Specify the folder that contains the file, and then select the file. Click
Open to recall the saved limit table on the screen.
6. You can recall a limit table from a trace on any channel
independently of the channel and trace that were active when
the limit table was saved to the file.
The ripple limit table is saved in the following format:
• In the first line, the channel number for the active channel at the
time of file saving will be output.
• In the second line, the trace number for the active trace at the time
of file saving will be output.
• The third line is a header indicating the segment items that are
output from the fourth line onward.
• From the fourth line onward, the segment data are output.
Operational procedure
"# Channel 1"
"# Trace 1"
Type, Begin Stimulus, End Stimulus, MaxRipple
ON, 933.0000000 MHz, 964.0000000 MHz, 1.5 dB
ON, 938.0000000 MHz, 953.0000000 MHz, 500 mdB
ON, 953.0000000 MHz, 960.0000000 MHz, 300 mdB
Turning ON/OFF Ripple Test and Result Display
You can set the limit test ON/OFF for each trace individually.
1. Press Channel Next/Channel Prev keys and Trace Next/Trace prev keys
to select the trace on which you want to apply the ripple test
function.
2. Press Analysis key to display the Analysis menu.
3. Click Ripple Limit to display the softkeys for to the ripple test.
4. Click the following corresponding softkeys.
Softkey Function
Ripple
Sets the ripple test ON/OFF.

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals