134
Enhanced Response Calibration
Overview
Procedure
Other topics about Basic Calibration
Overview
In enhanced response calibration, calibration data are measured by
connecting an OPEN standard, a SHORT standard, or a LOAD standard to
the output port (or a THRU standard between two ports).This calibration
effectively eliminates the directivity error, crosstalk, source match error,
frequency response reflection tracking error, and frequency response
transmission tracking error from the test setup in a transmission or
reflection test using those ports .
1. System Z0 should be changed to 1 ohm before calibration when
using waveguide calibration kit and measuring waveguide
devices.
2-Port Error Model (Enhanced Response)
Procedure
Connecting the Standard at Enhanced Response Calibration