Measurement
135
1. Press Channel Next/Channel Prev keys to select the channel for which
you want to perform the calibration.
2. Press Cal key.
3. Click Calibrate.
4. Click Enhanced Response.
5. Click Ports to select the test ports on which an enhanced response
calibration will be performed.
6. Connect an OPEN calibration standard to the output port.
7. Click Open to start the calibration measurement.
o If you select the calibration kit which has different
calibration definitions for each gender, (m) and (f) in the
name (label) of the standard displayed in the softkey
indicate male (m) and female (f) for the analyzer's
connector, respectively.
8. Disconnect the OPEN calibration standard and replace it with a
SHORT calibration standard.
9. Click Short to start the calibration measurement.
10. Disconnect the SHORT calibration standard and replace it with a
LOAD standard.
11. Click Load to start the calibration measurement.
12. Make a THRU connection between the two ports.
13. Click Thru to start the calibration measurement.
14. If an isolation calibration must be performed using a LOAD
standard, follow the procedure below:
13.
a. Connect a LOAD standard to the two test ports.

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals