Measurement
133
b. Click Isolation (Optional) to start the calibration measurement.
7. Click Done to terminate the response calibration (and the LOAD
isolation calibration) process. Upon pressing this key, calibration
coefficients will be calculated and saved. The error correction
function will also be automatically enabled.