Measurement
129
2. Press Cal key.
3. Click Calibrate.
4. Select OPEN or SHORT response calibration.
Softkey Function
Response
(Open)
Displays softkeys for performing an OPEN response
calibration (response calibration with an OPEN
standard)
Response
(Short)
Displays softkeys for performing a SHORT response
calibration (response calibration with a SHORT
standard)
5.
6. Click Select Port.
7. Select the test port upon which you will perform OPEN/SHORT
response calibration.
8. Connect a calibration standard (OPEN or SHORT) to the selected test
port (connector to which the DUT is to be connected).
9. Click Open or Short to start the calibration measurement.
10. If an isolation calibration must be performed using a LOAD
standard, follow the procedure below.
11. Connect a LOAD standard to the selected test port (connector to
which the DUT is to be connected).
12. Click Load (Optional) to start the measurement on the LOAD
standard.
13. Click Done to terminate the response calibration (and the LOAD
isolation calibration) process. Upon pressing this key, calibration
coefficients will be calculated and saved. The error correction
function will also be automatically enabled.
Connecting standards in OPEN/SHORT Response calibration

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals