Measurement
141
21. Click Port x-y Isol to start the calibration measurement (x and y
denote the port numbers to which the LOAD standard is connected).
22. Click Return.
23. Click Done to terminate the full 2-port calibration process. Upon
pressing this key, calibration coefficients will be calculated and
saved. The error correction function will also be automatically
enabled.
Connecting standards in full 2-port calibration