1. Press Channel Next/Channel Prev keys to select the channel for which
you want to perform the calibration.
2. Press Cal key.
3. Click Calibrate > 4-Port Cal.
Reflection
1. Click Reflection.
2. Connect an OPEN calibration standard to test port 1.
3. Click Port 1 Open to start the calibration measurement.
o If you select the calibration kit which has different
calibration definitions for each gender, (m) and (f) in the
name (label) of the standard displayed in the softkey
indicate male (m) and female (f) for the analyzer's
connector, respectively.
4. Disconnect the OPEN calibration standard and replace it with a
SHORT calibration standard.
5. Click Port 1 Short to start the calibration measurement.
6. Disconnect the SHORT calibration standard and replace it with a
LOAD standard.
7. Click Port 1 Load to start the calibration measurement.
8. Repeat the above procedure for ports 2, 3 and 4.
9. Click Return.
Transmission
1. Click Transmission.
2. Make a THRU connection between ports 1 and 2.
3. Click Port 1-2 Thru to start the calibration measurement.
4. Repeat above procedure ports 1 and 3.
5. Repeat the procedure on ports 1 and 4.
6. Repeat the procedure on ports 2 and 3.
7. Repeat the procedure on ports 2 and 4.
8. Repeat the procedure on ports 3 and 4.
9. Click Return.
10. If an isolation calibration must be performed using a LOAD
standard, follow the procedure below.
11. Click Isolation (Optional).