EasyManuals Logo

Agilent Technologies E5071C User Manual

Agilent Technologies E5071C
2685 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #166 background imageLoading...
Page #166 background image
Measurement
165
10. When using an adapter to the ECal, click Characterization and
then press the softkey corresponding to the characterization for the
adapter you are using.
11. Click Confidence Check.
12. Compare the data trace and the memory trace and verify
whether measurement is correct.
13. The following is the procedure for comparison when
simultaneously displaying the data trace and the memory trace.
a. Press Display key.
b. Click Display > Data & Mem.
c. Press Scale key.
d. Click Auto Scale.
e. Determine whether the difference between the traces is
acceptable. The difference should be read in terms of linear
values instead of dB error. If we consider the magnitude of the
linear error as compared to the dB delta, the value is very
small. So to evaluate the difference between the traces, a
linear error scale should be used for comparison instead of the
dB error scale.
14. For all of the parameters you want to verify, repeat the
procedure.
Connecting ECal module (for verification of S21)

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals