E5071C
182
14. Click Offset Line 1 Thru.
15. Make a LOAD connection on the selected test port with a offset
load (X-band Delay line).
16. Click X-band Delay Line.
17. Click Done.
18. Click Return twice.
19. Now, all Short, Open and Load should have check mark on their
left.
20. Click Done.
The sweep range setting and Thru Offset frequency range should match
the frequency range of the offset load. Else, the load process is not
executed, hence no check mark appear on the Load softkey. Similarly, if
the frequency range is extended, the load process is also not executed.
Refer to redefining a calibration kit to learn more about it.