E5071C
196
3. Click Modify Cal Kit > Import Cal Kit... to open the dialog box.
4. Specify a folder, enter a file name, and click Open.
Restoring Definition File of Calibration Kit
1. Press Cal key.
2. Click Cal Kit, then select a calibration kit.
3. Click Modify Cal Kit > Restore Cal Kit....
4. Click OK to restore the definition of the calibration kit selected by Cal
Kit to factory default settings.