(Recommended) is displayed. The display after the execution of the
omissible line/match measurement is 2-3 Line/Match, and the display for
the remaining omissible line/match measurement is 2-3 (Optional).
The softkeys (Fwd and Rvs) for forward measurement and reverse
measurement for the omissible line/match measurement do not change.
For the simplified 4-port TRL calibration, up to three thru/line
measurements and line/match measurements can be omitted, but omitting
all of them results in poor accuracy.