E5071C
234
The electrical delay is automatically measured and can be confirmed.
Press Cal > Port Extensions > Extension Port 1 {Value}
4. If there is an offset in the short circuit used to terminate the
Unknown Thru standard, remember to subtract its electrical
length from the measurement result.
To perform the unknown Thru Calibration, follow the procedure below:
1. Press Cal key, then click Calkit.
2. Select the desired calibration kit (e.g. 85033D or User Kit)
3. Press Cal > Modify Calkit > Define STDs > No Name > Label
4. Type a name for unknown Thru Cal standard e.g. {Unknown Thru1}
• By changing the label to a different name (e.g. Unknown
Thru1), the 'No Name' standard will adopt to the specified
new name.
5. Press Cal > Modify Calkit > Define STDs > {Defined Name} > STD Type
> Unknown Thru
6. Click Offset Delay, then enter the approximate offset delay value of
the unknown Thru standard.
7. Press Cal > Modify Calkit > Specify CLSs > Thru > Set All > {Defined
Name}
6.
1. Different Unknown Thru can be used for different ports.
7. Press Cal > Calibrate > n-Port Cal (n>1) > Transmission > {Defined
Name} to execute mechanical Unknown Thru calibration.
8. Complete Reflection, and Isolation (optional) calibration for n full
port Cal.

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals