E5071C
244
15. Click ECal & Done. The calibration coefficient will be calculated,
turning on the error correction function automatically.
• If the ECal module is not connected to the E5071C or power
calibration have not been done yet, the ECal & Done menu item
will not be not available. In addition, if the ECal module is not
connected to the test port of the calibration target, an error may
occur.