Programming
2755
SCPI.SYSTem.TSET.EXTernal
Object type
Property (Read-Write)
Syntax
SCPI.SYSTem.TSET.EXTernal = Param
Param = SCPI.SYSTem.TSET.EXTernal
Description
This command turns on/off the external connection test set, and selects
the setting of the test set mode. When the external connection test set is
on (MODE1/MODE2), only the S11, S21,or absolute value measurements
are available. The fixture simulator's balance measurement function and
frequency offset function are not available. When the external connection
test set is changed, the calibration coefficients are cleared, and the error
correction function is turned off.
Variable
Parameter Param
Description External test set mode
Data type Character string type (String)
Range
Select from either of the following:
• "OFF": Turns off the external test set.
• "MODE1": Turns on the external test set, and sets
the test set mode to MODE1.
• "MODE2": Turns on the external test set, and sets
the test set mode to MODE2.
Preset
value
"OFF"
Examples
Dim ExtTset As String
SCPI.SYSTem.TSET.EXTernal = "MODE1"
ExtTset = SCPI.SYSTem.TSET.EXTernal
Equivalent key
System > Service Menu > External Test Set
Equivalent SCPI command