E5071C
2762
SCPI.TRIGger.OUTPut.STATe
Object type
Property (Read-Write)
Syntax
SCPI.TRIGger.OUTPut.STATe=Param
Param=SCPI.TRIGger.OUTPut.STATe
Description
This command sets/gets the External Trigger Output Port state.
Variable
Parameter Param
Description
Sets/Gets external trigger output
state.
Data type Boolean
Range
Select from either of the following:
True or ON: Turns ON the
External Trigger Output
False or OFF : Turns OFF the
External Trigger Output
Preset
value
"OFF"
Examples
Dim TrigStat As boolean
TrigStat=1
SCPI.TRIGger.OUTPut.STATe=TrigStat
Related objects
SCPI.TRIGger.OUTPut.POLarity
SCPI.TRIGger.OUTPut.POSition
Equivalent key
Trigger > Ext Trig Output
Equivalent SCPI command
Syntax
:TRIGger:OUTPut[:STATe] {ON|1|OFF|0}