E5071C
2768
SCPI.TRIGger.SEQuence.EXTernal.LLATency.STATe
Type of object
Property (Read-Write)
Syntax
SCPI.TRIGger.SEQuence.EXTernal.LLATency.STATe = Status
Status = SCPI.TRIGger.SEQuence.EXTernal.LLATency.STATe
Description
This command turns ON/OFF or returns the status of the low-latency
external trigger feature.
When turning on the low-latency external trigger feature, the point trigger
feature must be set to on and the trigger source must be set to external
trigger.
When the point trigger feature is set to off, or when the trigger
source is set to one other than the external trigger, the change
is ignored.
Variable
Parameter Status
Description On/off of the low-latency external trigger
Data type Boolean type (Boolean)
Range
Select from either of the following:
True or ON: Turns on the low-latency external
trigger.
False or OFF: Turns off the low-latency external
trigger.
Preset
value
False or OFF
Example of use
Dim Llat As Boolean
SCPI.TRIGger.SEQuence.EXTernal.LLATency.STATe = True
Llat = SCPI.TRIGger.SEQuence.EXTernal.LLATency.STATe
Related objects
SCPI.TRIGger.SEQuence.EXTernal.DELay
SCPI.TRIGger.SEQuence.POINt