EasyManua.ls Logo

Agilent Technologies E5071C - Page 2610

Agilent Technologies E5071C
2685 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming
2837
:CALCulate:TRACe{Tr}:TTIMe:STATe
Type
Command/Query
Syntax
:CALCulate:TRACe{Tr}:TTIMe:STATe {ON|OFF|1|0}
:CALCulate:TRACe{Tr}:TTIMe:STATe?
Description
This command displays the rise time marker.
Parameter
Description Status of display of rise time marker.
Data Type Boolean type (Boolean)
Range ON|OFF|1|0
Preset Value
OFF
Unit -
Resolution -
Query Response
{1|0} <newline><^END>
Examples
10 OUTPUT 717;":CALC:TRAC1:TTIM:STAT ON"
20 OUTPUT 717;":CALC:TRAC1:TTIM:STAT?"
30 ENTER 717;A
Related Commands
:CALCulate:TRACe{Tr}:TTIMe:DATA
:CALCulate:TRACe{Tr}:TTIMe:THReshold
Equivalent GUI
Marker Search > Rise Time (10-90%) or Rise Time (20-80%)