E5071C
262
(S21)
2.Use this to perform forward-direction measurement
only.
1-2 Rvs
(S12)
Executes line/match calibration for test ports 1 and
2.Use this to perform reverse-direction measurement
only.
1-3
Line/Match
Executes line/match calibration for test ports 1 and
3.Use this to perform forward-direction measurement
and reverse-direction measurement at the same time.
1-3 Fwd
(S31)
Executes line/match calibration for test ports 1 and
3.Use this to perform forward-direction measurement
only.
1-3 Rvs
(S13)
Executes line/match calibration for test ports 1 and
3.Use this to perform reverse-direction measurement
only.
1-4
Line/Match
Executes line/match calibration for test ports 1 and
4.Use this to perform forward-direction measurement
and reverse-direction measurement at the same time.
1-4 Fwd
(S41)
Executes line/match calibration for test ports 1 and
4.Use this to perform forward-direction measurement
only.
1-4 Rvs
(S14)
Executes line/match calibration for test ports 1 and
4.Use this to perform reverse-direction measurement
only.
2-3
Line/Match
Executes line/match calibration for test ports 2 and
3.Use this to perform forward-direction measurement
and reverse-direction measurement at the same time.
2-3 Fwd
(S32)
Executes line/match calibration for test ports 2 and
3.Use this to perform forward-direction measurement
only.
2-3 Rvs
(S23)
Executes line/match calibration for test ports 2 and
3.Use this to perform reverse-direction measurement
only.
2-4
Line/Match
Executes line/match calibration for test ports 2 and
4.Use this to perform forward-direction measurement